1 / 3
It is widely used in experiments with small sample volume, many steps, large centrifugal force, and low temperature requirements. Such as: Separation of samples in laboratories of biochemistry, genetic engineering, agriculture and forestry, food hygiene, hospitals and other scientific research institutions.
| Parameter | Specification |
|---|---|
| Model | CYKY-20D Desktop Multi-purpose High-speed Centrifuge |
| Max speed | 20586r/min |
| Max RCF | 29326×g |
| Timing Range | 0 ~ 99min59s |
| Speed Accuracy | ±10r/min |
| Motor | Frequency Conversion Motor |
| Power | AC 220V 50Hz 15A |
| Noise | ≤55dB |
| Weight (without rotor) | 37Kg |
| Dimensions | 400×495×378mm(L×W×H) |
| No. | Rotor Type | Max RPM | Max RCF (xg) | Capacity | Acc/Dec Time |
|---|---|---|---|---|---|
| NO.1 | Angle Rotor | 20586 | 29326 | 12x1.5ml | 10S / 30S |
| NO.2 | Angle Rotor | 16500 | 26937 | 24×1.5/2.2ml | 30S / 50S |
| NO.3 | Angle Rotor | 14000 | 19721 | 36x1.5ml | 40S / 50S |
| NO.4 | Angle Rotor | 13000 | 16626 | 48x1.5ml | 50S / 60S |
| NO.8 | Angle Rotor | 12000 | 16823 | 6×50ml | 60S / 60S |
| NO.10 | Angle Rotor | 10000 | 9690 | 4×100ml | 60S / 80S |